    Чертеж

    ULTRAMAT/OXYMAT 6, 19“ unit, dimensions in mm

    Схема подключения

    Pin assignment (electrical and gas connections)

    ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment

    ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment of Autocal board and PROFIBUS connectors

    ULTRAMAT/OXYMAT 6, 19“ unit, gas and electrical connections

    Технические данные

    General

    Position of use

    Front panel vertical

    Conformity

    CE identification EN 50081-1 and EN 50082-2

    Design, enclosure

    Weight

    Approx. 21 kg

    Degree of protection

    IP20 according to EN 60529

    Electrical characteristics

    EMC interference immunity
    (ElectroMagnetic Compatibility)

    According to standard requirements of NAMUR NE21 (08/98)

    Electrical safety

    According to EN 61010-1, overvoltage category III

    Power supply

    100 … 120 V AC (rated range 90 V … 132 V), 48 … 63 Hz or 200 … 240 V AC (rated range 180 V … 264 V), 48 … 63 Hz

    Power consumption

    approx. 70 VA

    Fuses

    120 ... 120 V: F1/F2 = T 1.6 A;

    200 ... 240 V: F1/F2 = T 1 A

    Electrical inputs and outputs (per channel)

    Analog output

    0/2/4 … 20 mA, floating, max. load 750 Ώ

    Relay outputs

    6, with changeover contacts, freely selectable, e.g. for range identification;

    loading capacity: 24 V AC/DC/1 A, floating, non sparking

    Analog inputs

    2, designed for 0/2/4 … 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference)

    Binary inputs

    6, designed for 24 V, floating, freely selectable, e.g. for range switching

    Serial interface

    RS 485

    Options

    Autocal function with 8 additional binary inputs and 8 relay outputs; also with PROFIBUS PA and PROFIBUS DP

    Ambient conditions

    Perm. ambient temperature

    -30 … +70 °C during storage and transport,

    +5 … +45 ºC during operation

    Permissible humidity

    < 90% relative humidity, during storage and transport (dew point must not be fallen below)



    Technical data, ULTRAMAT channel

     

    Measuring ranges

    4, switchable internally and externally; autoranging is also possible

    Smallest possible measuring range

    Depending on application, e.g.:

    CO: 0 … 10 vpm

    CO2: 0 … 5 vpm

    Largest possible measuring range

    Depending on application

    Measuring ranges with suppressed zero

    Any zero point is possible between 0 and 100%; smallest possible span 20%

    Characteristic

    Linearized

    Gas inlet conditions

    Permissible sample gas pressure

     
    • Without pressure switch

    600 … 1500 hPa (absolute)

    • With pressure switch

    600 … 1300 hPa (absolute)

    Sample gas flow

    18 … 90 l/h (0.3 … 1.5 l/min)

    Sample gas temperature

    0 … 50 ºC

    Sample gas humidity

    < 90 % relative humidity or depending on application, without condensation

    Time response

    Warm-up period

    With amb. temperature < 30 min (maximum accuracy achieved after 2 hours)

    Response time (T90 time)

    Dependent on length of analyzer cell, sample gas line and parameterizable damping

    Damping (electric time constant)

    0 to 100 s, programmable

    Dead time (purging time of gas path in analyzer at 1 l/min)

    Approx. 0.5 to 5 s, depending on version

    Time for internal signal processing

    < 1 s

    Pressure correction range

    Pressure sensor

     
    • internal

    600 … 1200 hPa absolute

    • external

    600 … 1500 hPa absolute

    Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)

    Output signal fluctuation

    ± 0,1% ... ± 1% of smallest possible measuring range specified on rating plate depending on the unit electronic time constant (corresponds to ± 0.33% with 2σ)

    Zero drift

    < 1% of measuring range/week

    Measured-value drift

    < 1% of measuring range/week

    Repeatability

    ≤ 1% of respective measuring range

    Linearity error

    < 0.5% of full-scale value

    Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)

    Ambient temperature

    < 1% of measuring range/10 K (only with stable EK temperature)

    Sample gas pressure

    With pressure compensation: < 0.15% of span/1% change in atmospheric pressure

    without pressure compensation: < 1.5% of span/1% change in atmospheric pressure

    Sample gas flow

    Negligible

    Power supply

    < 0.1% of output signal span with rated voltage ± 10%

    Ambient conditions

    Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases



    Technical data, OXYMAT channel

    Measuring ranges

    4, switchable internally and externally; autoranging is also possible

    Smallest possible measuring span (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)

    0.5% v/v, 2% v/v or 5% v/v O2

    Largest possible measuring span

    100 % v/v O2

    Measuring ranges with suppressed zero

    Any zero point is possible between 0 and 100% v/v as long as a suitable reference gas is used

    Gas inlet conditions

     

    Permissible sample gas pressure

     
    • for analyzers with pipes

    500 … 3000 hPa absolute

    • for analyzers with hoses
     
    • without pressure switch

    500 … 1500 hPa absolute

    • with pressure switch

    500 … 1300 hPa absolute

    Sample gas flow

    18 … 60 l/h (0.3 … 1 l/min)

    Sample gas temperature

    0 … 50 ºC

    Sample gas humidity

    < 90% RH (relative humidity)

    Time response

    Warm-up period

    With ambient temperature < 30 min (maximum accuracy achieved after 2 hours)

    Reading delay time

    min. 1.5 … 3.5 s, depending on version

    Damping (electric time constant)

    0 to 100 s, programmable

    Dead time (purging time of gas path in analyzer at 1 l/min)

    Approx. 0.5 … 2.5 s depending on version

    Time for internal signal processing

    < 1 s

    Pressure correction range

    Pressure sensor

     
    • internal

    500 … 2000 hPa absolute

    • external

    500 … 3000 hPa absolute

    Measuring response (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)

    Output signal fluctuation

    < 0.75% of smallest possible measuring range specified on rating plate with an electronic time constant of 1 s (corresponds to ± 0.25% with 2σ)

    Zero drift

    < 0.5%/month of smallest possible meas. span specified on rating plate

    Measured-value drift

    < 0.5%/month of respective measuring span

    Repeatability

    < 1%/month of respective measuring span

    Linearity error

    < 1%/month of respective measuring span

    Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)

    Ambient temperature

    < 0.5%/10 K referred to the smallest possible measuring span according to rating plate with span 0.5%: 1%/10 K

    Sample gas pressure (with air (100 hPa) as reference gas, a correction of the atmospheric pressure fluctuations is only possible when the sample gas is vented to ambient air)

    Without pressure compensation: < 2% of measuring span/1% change in pressure;

    with pressure compensation: < 0.2% of measuring span/1% change in pressure

    Residual gases

    Deviation in zero point corresponding to paramagnetic or diamagnetic deviation of residual gas

    Sample gas flow

    < 1% of smallest possible measuring span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range

    Power supply

    < 0.1% of output signal span with rated voltage ± 10%